Skip to content

Unverified

FS10A/FS20A

by Hunan Accurate Bio-Medical Technology Co., Ltd.

new-tab-link Click here to see more information.

Fingertip

Type
single-sqi-icon

YES

Signal quality indicator
waveform-icon

Waveform

indicator

OpenOx Performance

Root mean square error (ARMS) is a common measure of pulse oximeter device performance that combines bias and precision. Here we report Arms based on Open Oximetry device testing using 2013 FDA Guidelines for 510k submissions and 2017 ISO 80601, while also trying to account for expanded criteria to improve diversity of skin pigment in study cohorts (US FDA "Approach for Improving the Performance Evaluation of Pulse Oximeter Devices Taking Into Consideration Skin Pigmentation, Race and Ethnicity"). Read more about quantification of oximeter performance on our FAQ.

*NOTE: performance is only reported here once we have tested the device in ≥10 study subjects (i.e. as required by 2013 FDA and 2017 ISO requirements). Performance may change significantly as we continue to perform testing in additional subjects and conditions. Please continue to check back as we update frequently. Click the device to see details on how many subjects have been tested as well as details of skin color testing.

No data

Purchase Cost

Here we report retail purchase costs (USD) for buying the pulse oximeter, including one adult finger probe. Costs are obtained from one or multiple sources including manufacturers or online retail stores. Of note, some devices have special discount pricing for low and middle-income countries. The special prices are not accounted for in this report.

20

Lifetime Cost

Here we estimate the 10-year lifetime cost of ownership for this type of pulse oximeter (Caution: We make many assumptions!). Click the settings button next to the cost to see the formula and adjust these assumptions to your local data.

Specification Overview

Model

FS10A/FS20A

Alias Rebranding

Some oximeters may be rebranded, relabeled and sold by multiple distributors under different names. We are attempting to compile 'aliases' for devices.

Ulaif FS 20E

Type

We categorize devices as fingertip, handheld, tabletop, multiparameter, phone-based or wearable.

Fingertip

Reflectance or Transmittance

Some devices may have the capability to function with transmission or reflectance probes. Read more about the difference between reflectance and transmission devices in our FAQ.

Transmittance

Patient population

This indicates the intended patient populations for the device (adult, pediatrics, neonates), as specified by our review of the manufacturers' published specifications. Use in certain patient populations may require procurement of a separate probe.

Adult

Where made

This indicates the location where the device is manufactured as stated by the manufacturer (or the stated location of the manufacturer). Please note, devices may contain components manufactured in different location.

China

Ingress Protection (IP)

"Ingress Protection" ratings define levels of sealing effectiveness of devices from foreign bodies (e.g. dust) and moisture. Read more at our FAQ.

IPX2

Cost

Here we report retail purchase costs (USD) for buying the pulse oximeter, including one adult finger probe. Costs are obtained from one or multiple sources including manufacturers or online retail stores. Of note, some devices have special discount pricing for low and middle-income countries. The special prices are not accounted for in this report.

$20

Lifetime Cost

Here we estimate the 10-year lifetime cost of ownership for this type of pulse oximeter (Caution: We make many assumptions!). Click the settings button next to the cost to see the formula and adjust these assumptions to your local data.

760.00

Features

Here we report device features such as signal quality indicator, waveform, carboxy-Hb, perfusion index and ability to measure Hb. These are based on review of manufacturers' manuals and may be incomplete.

Extended low perfusion performance testing

Standard Performance info

Manufacturer claimed Arms (root mean square error) for SpO2 70-100%

Here we report the root mean square error (ARMS) as provided in the manufacturer’s product manual or other literature, which may include data from the 510(k) submission.

1.1-2%

Independent Arms (root mean square error) for SpO2 70-100%

Root mean square error (ARMS) is a common measure of pulse oximeter device performance that combines bias and precision. Here we report Arms based on Open Oximetry device testing using 2013 FDA Guidelines for 510k submissions and 2017 ISO 80601, while also trying to account for expanded criteria to improve diversity of skin pigment in study cohorts (US FDA "Approach for Improving the Performance Evaluation of Pulse Oximeter Devices Taking Into Consideration Skin Pigmentation, Race and Ethnicity"). Read more about quantification of oximeter performance on our FAQ.

*NOTE: performance is only reported here once we have tested the device in ≥10 study subjects (i.e. as required by 2013 FDA and 2017 ISO requirements). Performance may change significantly as we continue to perform testing in additional subjects and conditions. Please continue to check back as we update frequently. Click the device to see details on how many subjects have been tested as well as details of skin color testing.

No data

Source of independent Arms data

Root mean square error (ARMS) is a common measure of pulse oximeter device performance. 'ARMS' may be ascertained from manufacturers' published data, 510k reports, package inserts or primary data from testing conducted by the UCSF Hypoxia Lab. Devices independently tested by the Open Oximetry Project will be marked 'verified' or 'failed' depending on study findings.

Manufacturer spec sheet

510k

FDA 510k approval. Read more about FDA and 510k approvals on our FAQ.

510k date

Date FDA 510k approved

12/04/2014

CE-XXXX

Open Oximetry attempts to request CE certificates from manufacturers and distributors though this is not always possible. CE numbers shown here are largely obtained from manufacturer's literature and are unverified by our team. Read more about CE marking in our FAQ..

CE-0123

Extended Performance info

Peer reviewed Manuscripts

Here are select manuscripts published in the peer-reviewed literature.

Data published by manufacturer

We are working to gather device performance data published directly by manufacturers.

Perfusion performance data

We collected the percent modulation of the infrared (IR) signal—often termed pulsatility amplitude or perfusion index (PI)—from our reference devices, Nellcor PM1000N and Masimo Rad-97. This measure serves as an indirect, albeit imperfect, surrogate for perfusion and signal strength.

This column provides insights into the pulsatility amplitude distribution of the cohort by reporting the median pulsatility amplitude and interquartile range (Q1–Q3). The median pulsatility amplitude represents the typical blood flow strength of the collected samples, while the interquartile range (Q1-Q3) shows the middle 50% of pulsatility amplitude values, giving an idea of how much variation exists around this typical flow level.

We report pulsatility amplitude from the Nellcor PM1000N for all devices, except for the Masimo Rad-97, which provides its own PI measurement. For consistency across devices, we applied a correction factor to the PM1000N pulsatility amplitude by dividing the values by 10, making them comparable to the pulsatility amplitude reported by other devices, such as Masimo.

Not Available

Raw PPG data

We are working to gather raw data for device performance to share for independent analysis. We expect to launch this feature soon.

Not Available

In vitro (simulator) performance data

We are working on novel in vitro testing protocols for both commercially available devices (e.g. Fluke ProSim8) and novel in vitro devices. We expect to report data for this testing soon.

Not Available

Real world clinical data

Here we link to studies conducted in the clinical settings.

Not Available

Submit a question