(2013) Pulse Oximeter Model Ab-F1/Ah-S1/Ah-M1

by ACARE TECHNOLOGY CO., LTD.

About this oximeter:

Acare has made several oximeter models. The Hypoxia Lab has published data on the 2013 AH-MX which performed to meet FDA standards with Arms < 3%. The Hypoxia Lab has also tested the newer generations of AH-M1 2017 which performed with Arms of 3.67, and the AH-M1 2020 performed with an Arms of 4.89.

Click here to see more information.

Handheld

Type

2.1-3%

Arms

YES

Signal quality indicator

Specification Overview

Manufacturer

ACARE TECHNOLOGY CO., LTD.

Model

(2013) Pulse Oximeter Model Ab-F1/Ah-S1/Ah-M1

Alias Rebranding

Lifebox

Type

Handheld

Reflectance or Transmittance

Transmittance

Patient population

Adult, Pediatrics, Neonates

Where made

Taiwan

Features

Perfusion Index

Performance info

Arms (root mean square error) for SpO2 70-100%

2.1-3%

Source of Arms data

UCSF Hypoxia Lab

510k

510k date

12/22/11

CE-XXXX

CE -

Peer reviewed Manuscripts

Date tested by Hypoxia Lab

2013

Submit a question
About Oximeter Database Oximetry FAQ Study Protocols Publications